Meatest M142 Multifunction Calibrator 10ppm (10 parts per million) Accuracy

Code: M142

10 ppm AC/DC multifunction calibrator with built-in multimeter, 30 A extended current range and variety of functions to calibrate most of the instruments your electrical laboratory will ever come across with ease.

  • 10 ppm accuracy
  • 30 A range, variety of functions
  • Built-in multimeter

2,489,990.00 1,989,990.00

In Stock 2-3 Weeks

In Stock 2-3 Weeks

Description

The M142 is calibrator of electric quantities for application in calibration laboratories and in production lines where voltage, current, resistance, capacity and frequency meters are manufactured. Load capacity of the voltage output is 30 mA – enough for most high-consumption analogue power-meters. Installed harmonic and non-harmonic shape signals allow for testing meter sensitivity to distorted signals by a signal with various crest factor. Frequency modes, suitable for calibration of multimeters and time bases of oscilloscopes, have adjustable 6-digit frequency, amplitude and duty ratio of the output signal. The calibrator can measure temperature with TC and RTD temperature sensors to show it on display or use for cold junction compensation.

Built-in process multimeter, standard part of M142 full version, can be used independently or simultaneously with source functions which makes testing transducers, regulators and evaluation units really easy. Using a single instrument you can evaluate output signals of various types of transducers and external sensors (strain gauge, pressure, torsion, strength, etc.), read them directly from calibrator display and use them in your calibrations.

M142 calibrator is focused on user friendliness making calibrations simple and straightforward. Each function has its dedicated chunky button and all relevant details including total relative accuracy are shown on large LCD. Range of adapters for easier terminal connection and automation possibilities, relative deviation feature that can help you with UUT error calculation and easy keyboard-controlled adjustment procedure are just a few examples to prove the point.

GPIB and RS-232 interfaces are present in both versions so the calibrator can be operated via remote control or used in WinQbase/Caliber software calibration systems including the CamOCR module for automated optical UUT readouts.

Standard score of delivery includes power cord, operation manual, test cables, factory calibration certificate, and RS23 cable.

Specifications

DC voltage 0 – 1000 V
± 10 ppm / year
AC voltage 0 – 1000 V, 20 Hz – 100 kHz
sinusoidal & non-sinusoidal waveforms
± 180 ppm / year
DC current 0 – 30 A (1500 A with 140-50 Current Coil)
± 100 ppm / year
AC current 1 uA – 30 A, 20 Hz – 10 kHz (1500 A with 140-50 Current Coil)
sinusoidal & non-sinusoidal waveforms
± 500 ppm / year
DC Power & Energy 0.4 mW – 4.8 kW (240 kW with 140-50 Current Coil)
1-1999 s
± 510 ppm / year
AC Power & Energy 0.4 mW – 4.8 kW, 40 – 400 Hz (240 kW with 140-50 Current Coil)
1-1999 s, power factor -1 to +1
± 635 ppm / year
Resistance 0 Ω – 1 GΩ, resolution 4½ digits
± 100 ppm / year
Capacitance 700 pF – 100 uF, resolution 4½ digits
± 0.5 % / year
Temperature TC BEJKNRST with cold junction compensation
RTD ± 0.07°C / year
Frequency 0.1 Hz – 20 MHz
± 50 ppm / year

Process Multimeter

DC voltage 0 – 20 V
mVDC resolution down to 100 nV
± 100 ppm / year
DC current 0 – 25 mA
± 150 ppm / year
Resistance 0 Ω – 2.5 kΩ, resolution 4½ digits
± 200 ppm / year
Temperature TC BEJKNRST
RTD ± 0.1°C / year
Frequency 1 Hz – 15 kHz
± 50 ppm / year

Downloads

Datasheet


M142 Datasheet

Language: English
File Size: 2.61 KB

Manual


M142 Manual

Language: English
File Size: 3.00 MB

Slide


M142 Slide

Language: English
File Size: 3.00 MB

Reviews

There are no reviews yet.

Be the first to review “Meatest M142 Multifunction Calibrator 10ppm (10 parts per million) Accuracy”

Your email address will not be published. Required fields are marked *