Automatic Double-Waveform Impulse

Model:LCG C Series

Automatic double-waveform impulse current testing system LCG C series are comprehensive impulse current testing systems with high system integration.

In Stock
Enquire Now

Description

LCG C series are comprehensive impulse current testing systems with high system integration, intelligent control, easy operation. Equipped with different modules, different waveforms can be generated, ideal for one-stop testing solutions. The automatic control system allows for easy test setting, waveform switching and one-button automatic running.

Features:

  • Open structure, 3D centripetal layout, with small inductance and even discharge;
  • Easy & flexible operation with highly integrated automation (test scheduling); automatic control system: MCS 2000;
  • High capacity: impulse current up to 30 kA for Level 1, up to 200 kA for Level 2; charging voltage up to 80 kV;
  • Two waveforms, free switching;
  • Optional LCM 25D module, to raise the output current (10 / 350 μs);

Standard

  • IEC 61643-1
  • GB11032
  • GB/T 18802.1
  • UL1449
  • YD-T 5098
  • TB/T 3074
  • YD/T 1235.2

Specifications

Technical Parameters
Current Waveform 8 / 20 μs 10 / 350 μs
Peak Output Current 65 kA, 120 kA, 150 kA, 200 kA 5 kA, 15 kA, 25 kA, 30 kA
Front Time T1 8μS ±10% <50 μs
Tail Time T2 20 μs ±10% 350 μs ±10%
Reverse Overshooting <20% N/A
Wave-front oscillatory <5% <5%
Voltage Waveform Distortion <5% (sine wave)
Voltage polarity +/-, (automatic switch)
Waveform switch Automatic switch
Charging time 10s~999s
Charging voltage
instability
Less than 1%
External sync input 50Hz
Trigger mode Automatic or manual, single trigger available
Trigger Error Rate Less than 1%
Trigger range 100% rated charging voltage
Phase test range 0~360°
Operating Duty Test Internal, 0~500V, 5A continuous supply
Display screen 10 inch TFT color touch screen
Measurement accuracy 3%
Type of EUT MOV, TVS, GDT, SPD, power supply type SPD etc
EUT residual voltage <5kV
Current measurement
(optional)
Built-in test box, Ross coil, BNC output;
Voltage measurement
(optional)
Built-in test box, pure voltage divider, BNC output
Oscilloscope MDO3012
Insulation strength ≤500V RMS(common port COMMON to ground reference PE)
General Parameters
Operating Power supply AC220V±10%,50/60Hz, 10 A
Operating Temperature -10-+40℃
Operating Humidity 45%-75%RH(no condensation)
Grounding Resistor ≤0.5Ω
Dimension of Test Room 5 x 5 m (main body) plus 3 x 5 m (control unit)
Dimension of Operating Platform 1,800 x 1,100 x 1,000 m
Option of Models
Model LCG 65C LCG 120C LCG 150C LCG 200C
8/20 Peak Value 65kA 120kA 150kA 200kA
10/350 Peak Value 5kA 10kA 25kA 30kA
Stored energy 20kJ 74kJ 128kJ 200kJ
8/20 Charging voltage 30kV 60kV 60kV 80kV
10/350 Charging voltage 30kV 60kV 60kV 80kV
Operating Power Supply AC220V 32A AC220V 50A AC220V 50A AC220V 100A
Weight 1500kG 2000kG 2500kG 3500kG
Other Waveforms
No. Waveform Type Max Output Charging Voltage Equivalent Output Impedance
A 10/350 Current waveform 50 kA 20 kV 0.4Ω
B 4/10 Peak Value Current waveform 100 kA 100 kV
C 8/20 Current waveform 250 kA 100 kV 0.4Ω
D 10/1000 Current waveform 20 kA 15 kV 0.5Ω
E 100/1000 Current waveform 20 kA 15 kV 0.5Ω
F 1/10 Current waveform 20 kA 50kV 2.5Ω
P Other waveforms can be customized
Note: The default configuration for each equipment of this series is equipped with two waveform modules. Other configuration can be customized.