Rigol PVP3150 – 150 MHz Passive Probe with 1x:10x
Frequency Compensation
Before taking any measurement using a probe, first check the frequency compensation of the probe and adjust it to match the oscilloscope used. Generally, the oscilloscope provides a calibration signal output terminal at the front panel. Connect the probe (set the probe attenuation to 10X) to the signal output terminal to display a square waveform on the oscilloscope. Insert the adjustment tool into the compensation adjustment hole at the probe head and rotate it clockwise/counterclockwise until the properly compensated square waveform is displayed (flat-top square waveform).
Undercompensated
Overcompensated
Properly Compensated
Manufacturer Specs
General Specifications | |
Electrical Specifications | |
Bandwidth | 1X: DC to 20 MHz 10X: DC to 150 MHz |
Rise Time | 1X: 17.5 ns 10X: 2.3 ns |
Attenuation Ratio | 10:1 or 1:1 Switchable |
Input Resistance | 1X: 1 MΩ ± 1% @ DC 10X: 10 MΩ ± 1% @ DC |
Input Capacitance | 1X: 50 pF ± 20 pF 10X: 10 pF ± 5 pF |
Max. Rated Input Voltage | 1X: 150 VRMS CAT II 10X: 300 VRMS CAT II |
Mechanical Characteristics | |
Weight (probe only) | About 45 g |
Cable Length | 120 ± 2 cm |
Probe Tip Diameter | 4.8 mm |
Environmental Specifications | |
Temperature | Operating: 0℃ to +50℃ Nonoperating: -20℃ to +70℃ |
Altitude | Operating: 2,000 m Nonoperating: 15,000 m |
Max. Relative Humidity | 80% relative humidity for temperature up to +31℃, decreasing linearly to 40% relative humidity at +50℃ |
Pollution Degree | Pollution Degree 2 |